Tube Furnace |
Structure |
-
Double layer steel casing with air cooling. Two cooling fans inside to keep the temperature of the outside case < 55°C.
-
High purity fibrous alumina insulation for Maximum energy saving.
|
Power Consumption |
5kW Max. |
Input Voltage |
AC 208~240V, Single Phase, 50/60Hz |
Max. Heating Temp. |
1700°C |
Max. Heating Rate |
5°C/min |
Max. Cooling Rate |
5°C/min |
Heating Zone Length |
455mm (18") |
Constant Temp. Zone |
150(+/-1C) |
Temperature Accuracy |
+/-1°C |
Temperature Controller |
-
PID automatic control via SCR (Silicon Controlled Rectifier) power control with phase angle fired, current limiting resistor.
-
30 programmable segments for precise control of heating rate, cooling rate and dwell time.
-
Built in PID Auto-Tune function with overheating & broken thermocouple protection.
-
Over temperature protection and alarm allows for operation without attendant(s).
-
+/- 1 ºC temperature accuracy.
-
RS485 Communications Port.
-
Advanced Temperature Control (Optional)
-
MTS02-Y Temperature control software kit (for YD518P series controllers) + 15" laptop package is available at the options bar.
-
You may upgrade the temperature controller to the Eurotherm 3504 Temperature Controller which is fully compatible with LabVIEW. This package includes a Eurotherm 3504 programmer, a communication cable (RS485 - USB) and a software CD.
|
Heating Elements |
1800 grade MoSi2 ( 8 pcs ) |
Tube Materials and Size |
-
99.8% high purity Al2O3 ceramic tube
-
ID: 73mm x OD: 80 mm x 1000 mm length.
|
Digital Vacuum Gauge |
-
Corrosion-proof Capacitance Diaphragm Gauge with ceramic coated sensor head specially designed for vacuum measurements with the presence of corrosive gasses.
-
Gauge is installed on the vacuum flange (right side) of tube furnace and provides a wide measure range (10^-5 ~ 1000 torr).
|
Vacuum Sealing |
-
Two stainless steel flanges with a digital vacuum gauge and valves are included.
-
Vacuum pressure can reach 5x10-2 torr via the mechanical pump.
-
The leaking rate is 6 mtorr/min in max. and 2 mtorr/min when stable, get details chart and data measured by CE certificated digital vacuum meter:
|
Vacuum Pump and fittings |
Vacuum Pump
|
-
Heavy Duty Rotary Vane Vacuum Pump (7.8 CFM -240 L/m) with Two Stage Exhaust System installed inside the vacuum station. vacuum pressure can reach up to 10^-2 Torr.
-
The max. vacuum pressure in the alumina tube furnace is 5x10^-2 torr by using this vacuum mechanical pump.
Option: For precise control of the vacuum pressure, we recommend you choose our turn-key solution EQ-VPC-MV. This is a vacuum pressure regulating system with motorized valve and gauge. It controls pressure range from 10^-6 torr ~ 760 torrs.
|
Fittings |
The standard vacuum fittings include:
-
A pair of stainless steel vacuum flange with KF-25 pump connector, two needle valves and one 1/4" Swagelok tube connector.
-
A 1000mm long KF-25 stainless steel vacuum bellows
-
Four KF-25 quick clamps
|
Gas Flow Meter |
-
Three gas flowmeters ( 4% accuracy FS) are installed inside mobile case provide a monitor to gas flowing rates.
-
1) Compact direct read flow meter, 10-100 cc/min.
-
2) Compact direct read flow meter, 16-160 cc/min.
-
3) Compact direct read flow meter, 25-250 cc/min.
-
One pressure meter is installed inside mobile cart for monitoring the pressure of the mixed gas.
-
Four stainless steel needle valve are installed in the front panel to control the gas route independently.
|
Oxygen Monitoring |
An Oxygen Sensor can be used to monitor the oxygen level of gas(es) used in CVD systems for preventing or reducing oxidation. |
Shipping Info |
Product Dimensions |
-
80 mm O.D Tube Furnace: 550 x 500 x 700mm L x W x H
-
Gas Mix Station: 600 x 600 x 700mm
-
Vacuum station: 600 x 600 x 700mm
|
New weight |
450 lb (in total) |
Shipping weight and Dimensions |
Three pallets: 220 lbs with 48x40x42", 200 lbs eith 39x31x40: and 180lbs with 39x31x40" dimension. |
Warranty |
One year limited warranty (Consumable parts such as processing tubes, O-rings, and heating elements
are not covered by the warranty, please order the replacement at related products page.) |